Remote Control Commands
R&S
®
NRX
324
User Manual 1178.5566.02 ─ 05
CALCulate<Measurement>[:CHANnel<Channel>]:ROSCillator:REFio:OUTPut[:
STATe]
<state>
Only effective for R&S
frequency selective power sensors.
If the REF connector of the power sensor is used as an output, enables or disables the
output signal.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<state>
*RST:
OFF
CALCulate<Measurement>[:CHANnel<Channel>]:ROSCillator:SOURce
<source>
Only effective for R&S
frequency selective power sensors.
Sets the source of the reference oscillator.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<source>
HOST | INTernal | REFio
*RST:
INTernal
[SENSe<Sensor>:]ADD
<sensor>
Adds a LAN power sensor and configures it automatically as long as the maximum
number of power sensors is not reached.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Setting parameters:
<sensor>
Hostname of the power sensor.
Example:
ADD "NRQ6-101435"
Usage:
Setting only
Manual operation:
See
Configuring Sensors