Remote Control Commands
R&S
®
NRX
317
User Manual 1178.5566.02 ─ 05
REPeat
Specifies that a measurement result is not output until the entire
measurement has been completed. This means that the number
of measurement cycle repetitions is equal to the set average
count. If the average count is large, the measurement time can
be very long.
The average count is set using
CHANnel<Channel>]:AVERage:COUNt[:VALue]
on page 332.
*RST:
MOVing
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:TYPE
<type>
Effective for continuous average, trace measurements.
Sets the averaging domain.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<type>
POWer | VIDeo | LINear
*RST:
POWer
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage[:STATe]
<value>
Effective for continuous average, burst average, timeslot measurements.
Enables the filter function of a power sensor. If enabled, the number of measured val-
ues is averaged. This reduces the effect of noise so that more reliable results are
obtained.
The number of measured values is set using
CHANnel<Channel>]:AVERage:COUNt[:VALue]
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
*RST:
ON
Manual operation:
See
Configuring Sensors