Remote Control Commands
R&S
®
NRX
310
User Manual 1178.5566.02 ─ 05
, compensates for the reflection of
the load at the signal output. Such a compensation is necessary if the voltage standing
wave ratio exceeds a value of 1.05.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<magnitude>
0.0
Disables the compensation.
Range:
-200.0 to 200.0
*RST:
0.0
Default unit: -
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet:STATe
<state>
Enables or disables the offset correction.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<state>
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet:TABLe:
INDex
<value>
Selects one of the available offset tables.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Setting parameters:
<value>
Range:
1 to 10
*RST:
1
Usage:
Setting only
Manual operation:
See
"Frequency dependent offset table"
Configuring Sensors