Remote Control Commands
R&S
®
NRX
231
User Manual 1178.5566.02 ─ 05
<Marker>
1 to 4
Marker (M1 to M4)
Parameters:
<time>
Range:
-15.0 to 15.0
*RST:
0.0
Default unit: s
Manual operation:
See
DISPlay[:WINDow<Window>]:TRACe:MARKer<Marker>:REFerence
<reference>
Sets one the available markers as reference.
Suffix:
<Window>
.
1 to 4
Measurement channel
<Marker>
1 to 4
Marker (M1 to M4)
Parameters:
<reference>
Range:
1 to 4
*RST:
1
Manual operation:
See
DISPlay[:WINDow<Window>]:TRACe:MARKer<Undef>:SELection
<markerNo>
Shows the selected marker in the trace.
Suffix:
<Window>
.
1 to 4
Measurement channel
<Undef>
1 to n
No suffix required.
Parameters:
<markerNo>
NONE | M1 | M2 | M3 | M4
*RST:
NONE
Manual operation:
See
14.4.7
Pulse Analysis
Further information:
●
[SENSe<Sensor>:]TRACe:MEASurement:OFFSet:TIME
................................................... 233
[SENSe<Sensor>:]TRACe:MID:OFFSet:TIME
...................................................................233
[SENSe<Sensor>:]TRACe:MID:TIME
...............................................................................233
CALCulate<Measurement>:TRACe:MEASurement:ALGorithm
........................................... 233
Measurement Settings and Results