Remote Control Commands
R&S
®
NRX
209
User Manual 1178.5566.02 ─ 05
Parameters:
<delay>
Range:
-5.0 to 10.0
*RST:
0.0
Default unit: s
Manual operation:
See
TRIGger<undef>:ALL:DTIMe
<dropout>
TRIGger<Measurement>[:CHANnel<Channel>]:DTIMe
<dropout>
Sets the dropout time for the internal trigger source. During this time, the signal power
must exceed (negative trigger slope) or undercut (positive trigger slope) the level
defined by the trigger level and trigger hysteresis. At least, this time must elapse before
triggering can occur again.
See
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<dropout>
Range:
0.0 to 10.0
*RST:
0.0
Default unit: s
Manual operation:
See
TRIGger<Measurement>[:CHANnel<Channel>]:EXTernal<Port>:IMPedance
<impedance>
Sets the termination resistance of the external trigger signal that is supplied at the trig-
ger input/output of the power sensor. Choose a setting that fits the impedance of the
trigger source to minimize reflections on the trigger signals.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
<Port>
1 to 2
Power sensor ports; 1 = USB port, 2 = trigger I/O connector
Parameters:
<impedance>
HIGH | LOW
Measurement Settings and Results