Remote Control Commands
R&S
®
NRX
187
User Manual 1178.5566.02 ─ 05
Parameters:
<value>
Range:
-1e18 to 1e18
*RST:
10.0
Default unit: -
Manual operation:
See
CALCulate<Measurement>:TRACe:Y[:SCALe]:TOP:WATT
<value>
Effective for trace measurements.
Specifies the reference value for the power axis.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<value>
Range:
-1e9 to 1e9
*RST:
1e-3
Default unit: W
Manual operation:
See
DISPlay[:WINDow<Window>]:ANALog:AUTO
<state>
DISPlay[:WINDow<Window>]:METer:AUTO
<state>
Automatically determines the scaling for the analog display. The upper and the lower
limit value are set depending on the current measurement data.
Suffix:
<Window>
.
1 to 4
Measurement channel
Parameters:
<state>
ONCE | OFF
*RST:
OFF
DISPlay[:WINDow<Window>]:ANALog:LOWer
<value>
DISPlay[:WINDow<Window>]:METer:LOWer
<value>
Sets the lower limit value of the analog scale.
Suffix:
<Window>
.
1 to 4
Measurement channel
Parameters:
<value>
Depends on the current output unit of the measured value.
Range:
1e-18 W to1e18 W; -150 DBM to 210 DBM; PCT:
1e-18 PCT to 1e22 PCT; -200 DB to 200 DB
Measurement Settings and Results