Remote Control Commands
R&S
®
NRX
183
User Manual 1178.5566.02 ─ 05
CALCulate<Measurement>:TRACe:X[:SCALe]:LEFT
<value>
Defines the position of the left screen edge relative to the delayed trigger. The value
may be negative so that signal components are displayed before the trigger event.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<value>
Range:
-15.0 to 15.0
*RST:
0.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>:TRACe:X[:SCALe]:LENGth
<value>
Sets the duration of the trace.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<value>
Range:
8.3e-9 to 30.0
*RST:
0.01
Default unit: s
Manual operation:
See
CALCulate<Measurement>:TRACe:Y[:SCALe]:SPAN:DB
<value>
Effective for trace measurements.
Specifies the value range for the power axis.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<value>
Range:
0.005 to 400.0
*RST:
50.0
Default unit: dB
Manual operation:
See
CALCulate<Measurement>:TRACe:Y[:SCALe]:SPAN:DBM
<value>
Effective for trace measurements.
Specifies the value range for the power axis.
Measurement Settings and Results