Remote Control Commands
R&S
®
NRX
181
User Manual 1178.5566.02 ─ 05
DISPlay:OVERload[:STATe]
<state>
If enabled, hides the sensor overload message. But be aware that overload can dam-
age the power sensor, depending on the amount of power and the duration of the over-
load condition.
You can query the allowed maximum power using
, or look it up in the data sheet of the power sensor.
Replaces the following R&S NRP2 command:
SERVice:DISPlay:OVERload
Parameters:
<state>
OFF | ON | NEVer
*RST:
ON; but does not apply if NEVer is set.
Manual operation:
See
"Hide Sensor Overload Message"
DISPlay:PIXMap?
Queries the display content. The return value is a binary block data, for example:
#577110xxxxxx...x
#577110
= block data header
xxxxxx...x
= binary format comprising an 8-bit BMP bitmap of the display content.
Usage:
Query only
DISPlay:UPDate
<mode>
Sets the update frequency of the measured values in the display.
Parameters:
<mode>
NORMal | SLOW | FREeze
FREeze
is useful if discontinuities in the voltage progress at the
analog outputs occur. In this state, the display does not con-
sume CPU time.
*RST:
NORMal
DISPlay[:WINDow<Window>][:STATe]
<state>
Opens or closes a measurement pane. See also
Chapter 5.1.5, "Selecting the Display
Suffix:
<Window>
.
1 to 4
Measurement channel
Parameters:
<state>
Measurement Settings and Results