Remote Control Commands
R&S
®
NRPxxS(N)
138
User Manual 1177.5079.02 ─ 10
SYSTem:RESTart
Restarts the firmware of the R&S NRPxxS(N) .
Usage:
Event
SYSTem:RUTime
<update_time>
Effective only in the NRP legacy mode.
Sets the result update time. That is the maximum rate in which the power sensor can
output measurement results. Relevant only in continuous measurement mode
(
ON).
Parameters:
<update_time>
Range:
0.0 to 10.0
*RST:
0.1
Default unit: Seconds
SYSTem:SERRor?
Queries the next static error, if available. Static errors are more severe than normal
error conditions, which can be queried using
Normal errors result from, for example, unknown commands or syntax errors and gen-
erally affect a single parameter or setting. Static errors, as a rule, prevent the execution
of normal measurements.
Static errors occur when you select conflicting settings. This could, for example, occur
in timeslot measurement mode with the following settings:
●
Width of a timeslot: 100
µs
●
Exclude time at the start of the slot: 40
µs
●
Exclude time at the end of the slot: 60
µs
Then there is "nothing left" to be measured, and a static error appears.
Usage:
Query only
SYSTem:SERRor:LIST:ALL?
Returns a list of all static errors that have occurred but have already been resolved.
For example, an overload of a short duration.
Example:
SYST:SERR:LIST:ALL?
Response:
0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification",0,"removed at uptime:2944;
notice; auto-averaging exceeded maximum time;
Notification"
.
Usage:
Query only