Instrument Functions
R&S
®
NRP2
73
User Manual 1173.9157.02 ─ 03
signal. Samples at equal time intervals over each acquisition interval are taken. From the
sum of the samples partial measurement results are formed. By forming the difference
from adjacent acquisitions, an improved result is achieved, that is, the DC offset voltage
and 1/f noise of the sensor amplifiers are significantly reduced. Depending on the
selected averaging number, a multiple of these results is averaged to further reduce
noise, see
chapter 4.2.3, "Filter / averaging"
The length of the acquisition interval depends on the set measurement mode:
Table 4-2: Acquisition interval of the respective measurement modes
Mode
Acquisition interval identical to ...
Cont Av
aperture time (default or user-specific)
T'slot Av
nominal width x number of slots
T'gate Av
gate length
Burst Av
burst length
After one acquisition has been completed, the output polarity of the detector output signal
is inverted. The next acquisition does not start until a sensor-specific settling time has
been passed. With the "T'slot Av", "T'gate Av" and "Burst Av" modes, where the start of
an acquisition must additionally be synchronized with the signal, the trigger period or a
multiple of it mainly determines the acquisition rate.
In "ContAv" mode, the acquisition interval can arbitrarily be set by defining the aperture
time. In most cases, the sensor-specific default value is sufficient for optimum perform-
ance. It is defined by the manufacturer as to effectively suppress 1/f noise. This means,
that for a given integration time (= averaging number x 2 x aperture time), the noise
content of the measurement result is minimal with the default setting.
Depending on the sensor, the default aperture time widely varies between 20 ms (for
R&S
NRP-Z11/21/22/23/24/31/32/28/98 multipath power sensors) and 10 µs (for
R&S
NRP-Z81/85/86 wideband power sensors). Thermal power sensor normally use a
5 ms default aperture time (20 ms in the past). Vastly differing averaging numbers when
using different sensor types are thus quite normal.
There are two cases where the default aperture time may not be the best selection:
●
if measuring signals with low-frequency modulation
●
if very short measurement times are needed, e.g. with remote-controlled applications
With low-frequency modulation, the aperture time should match the modulation period to
optimally suppress modulation-induced variations of the measurement result, see
ter 4.2.1.2, "Continuous average-power measurements"
Since this may lead to acquisitions that vastly exceed the 10 µs default aperture time of
wideband power sensors, a noticeable increase of fluctuations in the measurement result
due to noise must be considered with these sensors. Eventually the sensor-specific lower
measurement limit cannot be reached. The same applies for "T'slot Av", "T'gate Av" and
"Burst Av" power measurements when the acquisition rate drops down below 10 kHz for
these sensors. For details refer to the data sheet.
With R&S
NRP-Z11/21/22/23/24/31/32/28/98 multipath power sensors, a reduction of the
default aperture time may be essential when fast measurements - without averaging -
Data acquisition and parameters
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...