Programming Measurements
R&S
®
NRP2
381
User Manual 1173.9157.02 ─ 03
Example:
MEAS:ARRay? (1000)
1000 measured values are recorded in the ContAv mode. After termination of the last
measurement, the results are stored in the output queue. There is no triggering as in the
ContAv mode. The measurement result is a list of 1000 measured values separated by
a comma.
The keywords for processing two sensors can be added to the
MEAS:TSL
,
MEAS:BURS
and
MEAS:ARRay
commands, but the
:SWR
,
:RLOS
and
:REFL
functions are not avail-
able.
Example:
MEAS:TSL:RAT? 577 us, 8, 18 us, 18 us, DEF, 3, (@1), (@3)
The above timeslot example is extended to two sensors. The measurement result is a
list of 8 measured values separated by a comma for the 8 timeslots; these values were
obtained by division from the measured values of sensors 1 and 3.
C.1.5 Selecting the output unit for measured values
So far, no information has been given on the unit of the measured values. It is possible
to modify the output unit of the measured values. A distinction is made as to whether the
result represents a power or a power ratio. Relative measurements always deliver power
ratios. This also applies to the
:RATio
,
:SWR
,
:RLOS
and
:REFL
functions. The units
are set by means of the following commands:
UNIT:POWer DBM | W | DBUV
and
UNIT:RATio DB | DPCT | O
The default settings are
DBM
for power values and
DB
for power ratios.
Physical
unit
SCPI
notation
Meaning
dBm
DBM
Power in dB referenced to 1 mW: x / W = 10 log ( x / (1 mW) ) / dBm
Watt
W
dBµV
DBUV
Power in dB referred to (1µV)
2
/ 50 (: x dBm ( (x + 107) dBµV
dB
DB
Δ
%
DPCT
Deviation from 100% in %: x (% =
Δ
x + 100) %
1
O
When the
:SWR
,
:RLOS
and
:REFL
functions are used, the output unit, however, is
implicitly set by the
MEAS
command:
Measuring with high-level commands
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...