Remote Control - Commands
R&S
®
NRP2
268
User Manual 1173.9157.02 ─ 03
[SENSe<[1]..4>]:CORRection:SPDevice:SELect
selects one of these tables.
The
[SENSe<[1]..4>]:CORRection:SPDevice:STATe ON
command activates
measured-value correction.
Parameters:
<NR1>
int_value
*RST:
depending on sensor
[SENSe<[1]..4>]:CORRection:SPDevice:STATe
<state>
Instructs the sensor to perform a measured-value correction by means of one of the
stored S-parameter tables (S-parameter device).
Parameters:
<state>
ON | OFF
*RST:
depending on sensor
[SENSe<[1]..4>]:FREQuency[:CW]
<frequency>
[SENSe<[1]..4>]:FREQuency[:FIXed]
<frequency>
Informs the R&S
NRP2 about the frequency of the power to be measured since this
frequency is not automatically determined. The frequency is used to determine a fre-
quency-dependent correction factor for the measurement results.
Parameters:
<frequency>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: HZ
[SENSe<[1]..4>]:FUNCtion[:ON]
<mode>
Sets the sensor to the selected measurement mode.
In remote control, the timing (
SENS:TIM:EXCL:STAR
and -
:STOP)
and the trigger sys-
tem of all measurement modes (except the Burst mode) are user-selectable.
Note:
To increase measurement accuracy, the sensors measure each signal twice in the Burst,
Timeslot and Trace measurement modes (exception: in the Trace mode, with
SENS:TRAC:REAL ON
). As long as the trigger system is set to internal or external trig-
gering, this is usually of no importance to the user. If
TRIG:SOUR HOLD
or
BUS
is used,
however, then each measurement must be triggered twice (with
*TRG
or
TRIG:IMM
).
Error messages:
24
"Sensor mode not supported"
: A sensor does not support a measurement mode.
28
"Sensor not idle"
: The sensor is not in the IDLE state.
-151
"Invalid string data"
: An invalid parameter was transmitted for <string>.
SENSe (Sensor Commands)
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...