Instrument Functions
R&S
®
NRP2
156
User Manual 1173.9157.02 ─ 03
number of sensors connected and their measurement modes. With one sensor connec-
ted, operated in the ContAv mode, it is about 1 ms.
Switching on the averaging filter does not affect the refresh rate, but the speed at which
the output voltage changes can be expected to slow down as a result of the filter effect.
For applications having high requirements on the dynamic behavior of the analog outputs,
for example leveling loops, it is advisable to freeze the screen of the R&S
NRP2 to avoid
interruptions caused by writing to the display memory (
chapter 4.7.5, "Miscellaneous"
4.7.3 Sensor Check Source
The
Sensor Check Source
, option R&S NRP-B1, supplies an accurate 1 mW test signal
at 50 MHz. It can arbitrarily be switched on and off or be used in course of automatic test
measurements, see
chapter 4.7.6, "Test measurements"
► The signal is activated in the "System" menu.
Power Ref
Switches on/off the sensor check source.
SCPI command:
4.7.4 System information
This section describes how you retrieve information on the instrument like part number,
serial number, firmware version, etc. and installed options, as well as on connected power
sensors.
4.7.4.1
Information on sensors
► Select "Sensor Info..." in the "System" menu.
System settings
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...