Trace 1 / 2
Selects the trace (waveform) to be configured.
Up to two traces can be selected. This is also true if only one measurement channel
(sensor) is present. In this case, one trace for example can represent the average enve-
lope power and the other trace the peak or random power (available for the R&S NRP-
Z8x wideband sensors only).
Function
Selects the measurement function for the "Trace" mode. Only "Single" and "Ratio" func-
tions are available for trace measurements. Refer to
tion.
Note:
It is essential that the sensors of both channels support trace measurement, if the
"Ratio" function is selected.
Prim. / Sec. Channel
Selects the primary and secondary channel.
Eval.
Select the type of display. Since each display point (pixel) represents a time interval and
thus usually many samples, the R&S
NRP-Z8x wideband power sensors allow you to
select what is to be displayed.
"Average"
Average power value, resulting in a flicker-free display and smooth
trace.
Note:
Average is fixed for all multipath power sensors.
"Random"
Power of a randomly selected sample, i.e. a realistic display with signal
details.
"Peak"
Highest power measured (peak power).
Max. Hold
Displays the highest value measured for each point (pixel) of the selected display type
("Average", "Random ", "Peak").
"On / Off"
Starts/stops "Max Hold".
"Reset"
Resets the display and restarts recording.
4.5.4 Gates display mode
In the "Gate" mode, four separately configurable gates for signal evaluation are available.
The active gate is indicated in the window, the info panel shows the measurement results
and borders for this gate and the selected trace (waveform) . The settings for "Trig"gering,
"Filt"ering, "Function", and "Unit" are taken from the "Trace" display mode and cannot be
changed.
► Select "Gates" in the control panel opened via the Window
→
Graph
→
"Display" dialog
box ("Trace&Statistics" sensor mode).
Displaying traces
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...