Instrument Functions
R&S
®
NRP2
117
User Manual 1173.9157.02 ─ 03
Unit
Sets the result unit.
4.5.3.3
Filtering settings
You can smooth the trace in the "Average" display by means of an averaging filter. For
this purpose, a number of sweeps are recorded and the power of points of the same
position in time is averaged ("Trace Averaging"). The averaging process substantially
improves zero offset and linearity of the measurement result.
In the "Peak" display, a maximum search is performed over the number of curves corre-
sponding to the averaging number. If signals of stochastic envelope are applied, filtering
may increase the measurement accuracy. The fluctuations of results caused by the
inherent noise of the sensor will also be reduced, but this is unfavorable for a higher
systematic deviation (measured values are too high). Remedy: Reduce the video band-
width (see below).
The averaging filter does not affect the "Random" display.
To prevent results from being falsified due to instable trigger conditions, you should
always check whether the trace is stable in its time axis. This should be done without
averaging ("Trace Avg No = 1").
To meet the requirements for higher reproducibility and accuracy of numeric results,
measured power values for gates can be averaged even more by "Numeric Averaging",
irrespective of "Trace Averaging".
The measurement accuracy for the peak power, in particular for low power levels, can
be increased by reducing the effect of the inherent noise. You can do this by using a
smaller channel bandwidth ("Video Bw"). For the R&S
NRP-Z8x wideband power sen-
sors, this can be done in several steps between 300
kHz and 30
MHz (FULL). Reducing
the video bandwidth also increases the trigger sensitivity. As a rule of thumb, the trigger
threshold ("Level") can be reduced by 3 dB if the video bandwidth is reduced by a factor
of 4.
The video bandwidth must never be smaller than the RF bandwidth of the signal.
► Select "Filt..." in the trace control panel opened via the Window
→
Graph
→
"Display"
dialog box ("Trace&Statistics" sensor mode).
Displaying traces
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...