Remote Control - Commands 

R&S NRP-Z27/-Z37 

1169.3870.12 6.4 

E-2 

CALibration:ZERO:AUTO[?]   OFF | ON | ONCE 

The commands 

CALibration:ZERO:AUTO ON

 and 

CALibration:ZERO:AUTO ONCE

 perform zeroing. 

For this purpose, the test signal must be deactivated or the power sensor module disconnected from 
the signal source. The power sensor module automatically detects the presence of any significant 
power to be measured. This causes zeroing to be aborted and error message 

NRPERROR_CALZERO

 

to be output. The 

CALibration:ZERO:AUTO OFF

 is ignored. Zeroing takes four seconds at a minimum, 

but at least as long as the selected averaging filter needs for settling (only fixed-filter mode). 

 

Repeat zeroing 

during warm-up after switching on or connecting the instrument 

after a substantial variation of the ambient temperature 

after fastening the power sensor module to an RF connector at high temperature 

after several hours of operation 

when very low-power signals are to be measured, e.g. less than 10 dB above the 
lower measurement limit. 

For zeroing switch off the test signal and do not remove the power sensor module from 
the signal source. Apart from keeping the thermal balance, this has the advantage that the 
noise superimposed on the test signal (e.g. from a broadband amplifier) can be detected 
on zeroing and does not impair the measurement result. 

The query always yields 

1

(= 

OFF

). 

 

Default setting 

After a power-on reset, the zero offsets determined during the last calibration are used until the first 
zeroing. Therefore, very slight zero offsets are to be expected with a warmed up power sensor module. 
Initialization by means of 

*RST

 or 

SYSTem:INITialize

 has no influence on the current zero offsets. 

Содержание NRP-Z27

Страница 1: ...1169 3870 12 07 1 Operating Manual Power Sensor Module R S NRP Z27 1169 4102 02 R S NRP Z37 1169 3206 02 Test and Measurement...

Страница 2: ...1169 3870 12 07 2 Dear Customer R S is a registered trademark of Rohde Schwarz GmbH Co KG Trade names are trademarks of the owners...

Страница 3: ...rsonal injury or damage caused by dangerous situations Safety information is provided as follows In the Getting Started manual Chapter 1 1 Safety instructions The same information is provided in many...

Страница 4: ......

Страница 5: ...installed on the base unit must have version num ber 02 00 or later If you want to use the distortion compensation command group SENSe RGAMma see page 6 12 the Main Program must have at least version...

Страница 6: ......

Страница 7: ...2 1 5 Connecting the power sensor module 1 5 Connecting the power sensor module to the DUT 1 5 Performing measurements 1 5 PC control 1 6 Hardware and software requirements 1 6 Operation via the Passi...

Страница 8: ...e USB Adapter R S NRP Z3 1 7 Fig 1 5 Changing the primary adapter 1 8 Fig 1 6 Ports and LEDs on front panel 1 10 Fig 1 7 Ports on rear panel 1 10 Fig 1 8 Unlatching the R S Instrument connector 1 11 F...

Страница 9: ...ges To prevent this from happening never touch the inner conductor of the RF connector and never open the power sensor Connecting the sensor NOTICE To prevent EMI the power sensor module must never be...

Страница 10: ...Connecting the sensor R S NRP Z27 Z37 1169 3870 12 1 2 E 6 Do not apply excessive mechanical force to the RF connectors of the R S NRP Z27 Z37 and DUT F F F F...

Страница 11: ...in operation The multiple circular plug in connector must be inserted red marking upwards into the POWER SENSOR connector Then connect the 3 5 mm male connector of the RF cable with measurement input...

Страница 12: ...y To connect the power sensor module to the DUT more easily and without tilting it relieve it by slightly lifting it R S NRP Z27 Z37 R S FSMR POWER SENSOR RF INPUT DUT Fig 1 1 Configuration with the R...

Страница 13: ...commands in chapter 6 Connecting the power sensor module to the DUT The power sensor modules have a male N connector R S NRP Z27 or a male 3 5 mm connector R S NRP Z37 on their RF input and can thus...

Страница 14: ...nts are met the power sensor can be controlled using a suitable application program such as the NrpFlashup program contained in the Microsoft Windows version of the NRP Toolkit includes the modules Po...

Страница 15: ...and Scope modes The order in which the cables are connected is not critical Fig 1 4 Configuration with Active USB Adapter R S NRP Z3 The plug in power supply for the R S NRP Z3 can be powered from a...

Страница 16: ...circuit proof and has an internal fuse It is not possible to replace this fuse or open the plug in power supply NOTICE The plug in power supply is not intended for outdoor use Keep within the temperat...

Страница 17: ...ssive R S NRP Z4 interface adapter instead of a standard USB cable As an alternative to the supplied external power supply power can also be supplied using a DC voltage source with an output voltage o...

Страница 18: ...LEDs on front panel On the rear panel are the ports for power supply USB host trigger input and trigger output The ports on the rear panel are shown in Fig 1 7 1 Trigger output BNC socket 2 Trigger i...

Страница 19: ...orts and LEDs the R S NRP Z4 passive interface adapter can also be used as an alternative to a standard USB cable for connecting to the PC This adapter is connected to the R S Instrument port Simultan...

Страница 20: ...Connecting the sensor R S NRP Z27 Z37 1169 3870 12 1 12 E 6 Fig 1 9 Typical test setup with R S NRP Z5 and PC...

Страница 21: ...owing step by step description is based on an English version of Windows XP 1 Connect the R S NRP Z5 and put it into operation Connect at least one power sensor 2 Start the Device Manager and change t...

Страница 22: ...N or 2 92 mm 3 5 mm SMA connector respectively To connect the power sensor module to the DUT more easily and without tilting it relieve it by slightly lifting it Performing measurements The R S NRP u...

Страница 23: ...sors are generally connected to the USB ports of the test instrument via one of the two interface adapters R S NRP Z3 or R S NRP Z4 or the Sensor Hub R S NRP Z5 On some instruments such as the R S SMU...

Страница 24: ......

Страница 25: ...R S NRP Z27 Z37 Table of Contents Chapter 2 1169 3870 12 I 2 1 E 1 Table of Contents 2 Virtual Power Meter 2 1 Overview 2 1 Menus 2 3...

Страница 26: ...Figs and Tables Chapter 2 R S NRP Z27 Z37 1169 3870 12 I 2 2 E 1 Figs Fig 2 1 Power Viewer virtual power meter 2 1 Tables Table 2 1 Virtual power meter keys 2 2 Table 2 2 Virtual power meter entry fie...

Страница 27: ...after an extremely brief familiarization period you can perform measurements The other modules in NrpFlashup are described in Chapter 3 of the operating manual Terminal module or in the service manual...

Страница 28: ...o Turns auto averaging on or off When auto averaging is on the Length entry field has a grey background the current averaging factor is displayed Alt T Apply Accepts edited numerical values in the Fre...

Страница 29: ...file recording View Display Refresh Rate Opens a dialog box to adjust the display refresh rate The time in milliseconds between two refresh operations is entered The default setting is 200 ms Colours...

Страница 30: ...out the functions of the virtual power meter even without a power sensor module The display alternates between Measurement Value 1 Measurement Value 2 with a period given by Interval Simulation can be...

Страница 31: ...NRP Z27 Z37 Table of Contents Chapter 3 1169 3870 12 I 3 1 E 1 Table of Contents 3 Manual Operation 3 1 Program module Terminal 3 1 Main control elements 3 1 Menus 3 3 Program module Firmware Update...

Страница 32: ...1 Figs Fig 3 1 Sending commands using the Input field 3 2 Fig 3 2 Sending commands using command files 3 2 Tables Table 3 1 Buttons assigned to the Input field 3 2 Table 3 2 Buttons assigned to the Co...

Страница 33: ...e into the sensor The power splitter integrated in the power sensor module already occupies this data record Thus any adapters or attenuators connected ahead of the power sensor modules cannot be take...

Страница 34: ...ld to the power sensor module Alt S Loop With Loop the command or command sequence is cyclically sent Pressing the button again terminates the cyclic transmission The repetition rate is set in a dialo...

Страница 35: ...not forwarded to the power sensor module Measurement results parameters and status information that are returned are displayed in the Output field Table 3 3 Buttons assigned to the Output field Butto...

Страница 36: ...OK Response Time Opens the Response time dialog window where the response time of the power sensor module can be set Current indicates the time elapsed between dispatch of the last command and receipt...

Страница 37: ...mand sequences can be controlled In the Delay ms field the time interval for the cyclic transfer is specified in milliseconds The number of completed transfer cycles is displayed in the Counter field...

Страница 38: ...utomatically shifted upward and off the display if space is required for new values LF at EOT With this option active a line feed is appended to each response block from the sensor Delete on Start Wit...

Страница 39: ...R S NRP Z27 Z37 Program module Firmware Update 1169 3870 12 3 7 E 1 Program module Firmware Update A detailed description of the program module for firmware updates is provided in the Service Manual...

Страница 40: ......

Страница 41: ...R S NRP Z27 Z37 Table of Contents Chapter 5 1169 3870 12 I 5 1 E 1 Table of Contents 5 Remote Control Fundamentals 5 1...

Страница 42: ......

Страница 43: ...trol of R S NRP power sensors and power sensor modules This driver can be found on the CD ROM supplied with the sensor or downloaded in its most recent version via the internet http rohde schwarz com...

Страница 44: ......

Страница 45: ...SENSe AVERage COUNt AUTO RESolution 1 to 4 6 7 SENSe AVERage COUNt AUTO TYPE RESolution NSRatio 6 7 SENSe AVERage RESet 6 7 SENSe AVERage STATe OFF ON 6 8 SENSe AVERage TCONtrol MOVing REPeat 6 8 SENS...

Страница 46: ...6 16 TRIGger 6 17 ABORt 6 17 INITiate CONTinuous OFF ON 6 17 INITiate IMMediate 6 18 TRIGger COUNt 1 to 2 10 9 6 18 TRIGger DELay 0 to 100 0 6 18 TRIGger DELay AUTO OFF ON 6 18 TRIGger HOLDoff 0 0 to...

Страница 47: ...em 6 5 Table 6 3 Measurement mode POWer AVG 6 10 Table 6 4 Optimum selection of the sampling window size N 1 2 3 6 10 Table 6 5 Commands of the SYSTem system 6 13 Table 6 6 Meaning of Item in the SYST...

Страница 48: ......

Страница 49: ...COUNt is represented in the table as follows SENSe first level AVERage second level COUNt third level In the individual description the command is shown in full length An example of the command is giv...

Страница 50: ...g structure ROHDE SCHWARZ NRP Zx7 serial number firmware version serial number Serial number in ASCII firmware version Firmware version number in ASCII RST Reset RST sets the power sensor module to th...

Страница 51: ...th Bytes Query only ZERO AUTO OFF ON ONCE CALibration DATA calibration data set as definite length block CALibration DATA is used for writing a calibration data set in the flash memory of the power mo...

Страница 52: ...after a substantial variation of the ambient temperature after fastening the power sensor module to an RF connector at high temperature after several hours of operation when very low power signals ar...

Страница 53: ...arameter Unit Remarks SENSe AVERage RESet No query STATe OFF ON TCONtrol MOVing REPeat COUNt 1 to 65536 AUTO OFF ON ONCE TYPE RESolution NSRatio MTIMe 1 0 to 999 99 s NSRatio 0 0001 to 1 0 dB RESoluti...

Страница 54: ...h SENSe AVERage STATe ON so that the set averaging factor becomes effective Default setting 4 SENSe AVERage COUNt AUTO OFF ON ONCE SENSe AVERage COUNt AUTO activates auto averaging or deactivates fixe...

Страница 55: ...gher the selected index the better the measurement result is filtered without the last significant place 0 01 dB with an index of 3 actually being set The NSRatio setting is recommended instead The qu...

Страница 56: ...urement result This allows tendencies in the result to be recognized during the measurement procedure The REPeat parameter defines that a new result is output after the FIR filter has been filled with...

Страница 57: ...ctional coupler A negative offset can be used to correct the influence of a gain connected ahead The query yields the set offset in dB Default setting 0 0 dB SENSe CORRection OFFSet STATe OFF ON SENSe...

Страница 58: ...APERture defines the time interval sampling window measured values are continuously recorded in this interval In manual operation the default setting of 20 ms in conjunction with the activated smooth...

Страница 59: ...sed to suppress variations of results caused by low frequency modulation This parameter should be activated to reduce variations in results due to modulation when the size of the sampling window canno...

Страница 60: ...ex reflection coefficient of the source The query yields the set phase angle Default setting 0 0 SENSe RGAMma MAGNitude 0 0 to 1 0 With the SENSe RGAMma group of commands you can compensate the load d...

Страница 61: ...to IDN If no Item is specified the response string is a sequence of entries in the form Item Information String separated by CR and LF in C notation r n With the Item optionally appended to the comman...

Страница 62: ...has a nominal input impedance of 50 COUPLING DC The RF input of the power sensor modules is DC coupled CAL ABS date Date of absolute calibration in the format YYYY MM DD Invalid Calibration Date is re...

Страница 63: ...RANsaction BEGin marks the beginning of a sequence of setting commands between which the parameter limits must not be checked This prevents the display of error messages when a setting command causes...

Страница 64: ...triggers a selftest of the power sensor module In contrast to TST this command yields detailed information which is useful for troubleshooting No signal may be applied to the power sensor module whil...

Страница 65: ...dule enters the WAIT_FOR_TRIGGER state INITiate CONTinuous OFF ON INITiate CONTinuous ON activates the continuous measurement mode In this mode a new measurement is automatically started when a measur...

Страница 66: ...at the end of the measurement The TRIGger COUNt command does not define the number of trigger events required for performing the entire measurement task The number may vary depending on the measureme...

Страница 67: ...gger source The query yields the trigger hysteresis in dB Default setting 0 0 dB TRIGger IMMediate TRIGger IMMediate triggers a generic trigger event that causes the power sensor module to exit immedi...

Страница 68: ...ve 2 for NEGative Default setting POSitive TRIGger SOURce BUS EXTernal HOLD IMMediate INTernal TRIGger SOURce sets the trigger source BUS Triggering with command TRG or TRIGger IMMediate EXTernal Trig...

Страница 69: ...DATA LENGth Bytes 6 3 CALibration ZERO AUTO OFF ON ONCE OFF fixed 6 4 SENSe Commands SENSe AVERage COUNt 1 to 65536 4 6 6 SENSe AVERage COUNt AUTO OFF ON ONCE ON 6 6 SENSe AVERage COUNt AUTO MTIMe 1 0...

Страница 70: ...Se POWer AVG BUFFer STATe OFF ON OFF 6 11 SENSe POWer AVG SMOothing STATe OFF ON OFF 6 11 SENSe SGAMma CORRection STATe OFF ON OFF 6 12 SENSe SGAMma MAGNitude 0 0 to 1 0 0 0 6 12 SENSe SGAMma PHASe 36...

Страница 71: ...2 Command Parameter Unit Default setting Page TRIGger HYSTeresis 0 0 to 10 0 dB 0 0 6 19 TRIGger IMMediate 6 19 TRIGger LEVel x to y W 10 x 6 19 TRIGger SLOPe POSitive NEGative POSitive 6 20 TRIGger...

Страница 72: ...do not have the original packaging use similar materials that provide the same level of protection Securing When moving the product in a vehicle or using transporting equipment make sure that the pro...

Страница 73: ......

Страница 74: ...in mind that the casing is not waterproof If you use isopropyl alcohol or ethanol instead of water be careful not to damage the labeling Do not use cleaning agents that can damage the instrument such...

Страница 75: ...n disposal via the municipal collection points for waste electrical and electronic equipment is not permitted Labeling in line with EU directive WEEE Rohde Schwarz has developed a disposal concept for...

Отзывы: