![R&S NRP Series Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/rands/nrp-series/nrp-series_user-manual_774882057.webp)
Replacing an R&S
NRP
‑
Zxx with an R&S NRPxxA(N)
R&S
®
NRPxxA(N)
57
User Manual 1177.6017.02 ─ 08
8
Replacing an R&S
NRP
‑
Zxx with an R&S
NRPxxA(N)
The R&S NRPxxA(N) power sensors are compatible with the R&S
NRP
‑
Zxx series of
power sensors.
R&S
NRPxx power sensor
Replaces R&S
NRP
‑
Zxx power sensor
R&S
NRP6A
/ R&S
NRP6AN - USB connected
R&S
NRP-Z91
R&S
NRP18A
/ R&S
NRP18AN - USB connected
-
To use the new power sensors, it can be required to update the drivers. For computer-
based software applications (R&S
NRPV and R&S
Power Viewer), install the latest
R&S
NRP Toolkit (V
4.20 or higher).
For using the sensors with base units, signal generators, spectrum analyzers or other
Rohde & Schwarz instruments, install the latest firmware version.
8.1
Important difference
After powering the R&S NRPxxA(N), the firmware is loaded. Because the USB power
sensors are plug-and-play devices, the drivers are loaded after enumeration in the host
system. For the R&S
NRP
‑
Zxx power sensors, the whole process takes 4
s to 5
s. For
USB power sensors, the process can take up to 8
s.
Otherwise, the R&S
NRP
‑
Zxx power sensors and the USB power sensors are compati-
ble as far as possible.
8.2
Prerequisites
Install the R&S
NRP Toolkit V
4.20 or higher, see
The new version of the R&S
NRP Toolkit is compatible with both the R&S
NRP
‑
Zxx
and the R&S NRPxxA(N) so that its installation does not affect the usage of the
R&S
NRP
‑
Zxx.
After the new version of the R&S
NRP Toolkit is installed, you can connect the R&S
NRPxxA(N) to the computer and use it with Rohde & Schwarz software applications or
your own programs. For information on Rohde & Schwarz software applications, see
the release notes and the manual of the software application.
Prerequisites