User Manual 1177.6017.02 ─ 08
Magnitude ← Γ Correction
Sets the magnitude of the complex reflection coefficient of the source ,
Γ
source
.
A value of 0.0 corresponds to an ideal matched source. A value of 1.0 corresponds to
total reflection.
Remote command:
[SENSe<Sensor>:]SGAMma:MAGNitude
Phase ← Γ Correction
Sets the phase angle of the complex reflection coefficient of the source,
Γ
source
.
Remote command:
Zero Calibration
Performs zeroing using the signal at the sensor input. See
Note:
Turn off all test signals before zeroing. An active test signal during zeroing causes an
error.
Remote command:
CALibration<Channel>:ZERO:AUTO
Sensor Information
Displays information on the power sensor.
●
General information: manufacturer, sensor type, stock number, serial number, firm-
ware version
●
Network settings: MAC address, hostname, IP address, sensor name. A part of
them, you can configure under
Chapter 6.5.4, "System settings"
●
Important specifications: technology, frequency range, power limits, resolution, ...
●
Calibration information
Remote command:
6.5.2
Averaging settings
Describes the parameters for automatic averaging.
Further information:
●
Chapter 9.8.1, "Configuring auto averaging"
Access: main dialog of the web user interface > navigation pane > "Averaging"