Troubleshooting
R&S
®
NRPxxA(N)
149
User Manual 1177.6017.02 ─ 08
12.3
Problems during a firmware update
The firmware update is described in
Solutions for potential problems that can occur when using the Firmware Update for
NRP Family, see
Firmware update was interrupted
If for example, a power cut happened during the firmware update, problems can occur.
1. Perform the firmware update again. Sometimes, a further update fixes the prob-
lems.
2. If the power sensor is not accessible anymore, contact the service.
Firmware update was aborted
If there is not enough free storage space, the firmware update is aborted. An error
message is displayed, and the status LED of the power sensor starts flashing red.
1. Perform a sanitization procedure, as described in the instrument security proce-
dures. This document is available on the product page, see
on page 11, or the CD-ROM delivered with the power sensor.
2. Perform the firmware update again.
12.4
Cannot establish a LAN connection
If you have problems to establish a LAN connection as described in
on page 20, try the following measures:
●
Use the Configure Network Sensor component of the R&S
NRP Toolkit, see
●
"Troubleshooting for peer-to-peer connections"
12.5
Contacting customer support
Technical support – where and when you need it