Instrument Functions
R&S
®
NGL200/NGM200
55
User Manual 1178.8736.02 ─ 06
5. Activate the "Link to Channel 2" or "Link to Channel 1" menu item.
● ON: The linked channel is turned off when an OCP event is triggered.
● OFF: The linked channel is not affected when an OCP event is triggered.
6.5.2
Over Voltage Protection (OVP)
When the output voltage exceeds the limit set for the respective channel, an alert is
triggered and the affected channel is turned off according to the settings configured in
the OVP dialog.
1. Select "Over Voltage Protection (OVP)" from the menu.
The R&S
NGL/NGM displays the OVP dialog.
Figure 6-15: Over voltage protection dialog
2. Activate the "Enabled" menu item.
The R&S
NGL/NGM enables the OVP and displays the OVP icon on the selected
channel status bar information.
3. Set the required level for OVP.
The R&S
NGL/NGM displays the on-screen keypad to set the value.
4. Confirm value with the unit key (mV or V).
6.5.3
Over Power Protection (OPP)
When the output power exceeds the limit set for the respective channels, an alert is
triggered and the affected channels are turned off according to the settings configured
in the OPP dialog.
1. Select "Over Power Protection (OPP)" menu item.
The R&S
NGL/NGM displays the OPP dialog.
Protection