Remote Control Commands
R&S
®
NGL200
90
User Manual 1178.8736.02 ─ 02.01
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit[:UPPer]
<voltage>
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit[:UPPer]?
Sets/Queries the upper safety limit for voltage.
Setting parameters:
<voltage>
numeric | MIN | MAX
numeric
Numeric value for upper safety limit.
MIN
Min value for upper safety limit.
MAX
Max value for upper safety limit.
Range:
0.000E+00 to 2.000E+01
Increment: 0.001
*RST:
2.000E+01
Example:
See
Example "Configuring the output voltage"
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
<voltage>
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Sets/Queries the voltage value of the selected channel.
Parameters:
<voltage>
numeric | MIN | MAX | UP | DOWN
numeric
Numeric value in V.
MIN
Minimum voltage at 0.000 V.
MAX
Maximumm voltage at 20.000 V.
UP
Increase voltage by a defined step size. See
]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
DOWN
Decrease voltage by a defined step size. See
]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
on page 91.
Range:
0.000 to 20.00
Default unit: V
Example:
See
Example "Configuring the output voltage"
Configuration Commands