Remote Commands
R&S
®
FSW
942
User Manual 1173.9411.02 ─ 43
Example:
CALC:LIM:ESP:PCL1:MAX -40 dBm
Sets the maximum power value of the first power class to -40
dBm.
Manual operation:
See
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MINimum
<Level>
This command defines the lower limit of a particular power class.
Note:
●
The first power class always has a lower limit of -200
dBm.
●
The lower limit of a power class must always be the same as the upper limit of the
previous power class.
●
The power class must already exist (see
ESPectrum<sb>:PCLass<pc>:COUNt
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<sb>
1 to 8
Sub block in a Multi-SEM measurement
<pc>
1..n
power class
Parameters:
<Level>
Range:
-200 dBm to 199.9 dBm
Default unit: dBm
Example:
CALC:LIM:ESP:PCL2:MIN -40 dBm
Sets the minimum power value of the second power class to -40
dBm.
Manual operation:
See
14.5.7.7
Configuring MSR SEM Measurements
The following commands configure MSR SEM measurements. For details see
ter 7.6.4.4, "Multi-Standard Radio (MSR) SEM Measurements"
Chapter 7.6.5.5, "MSR Settings"
[SENSe:]ESPectrum<sb>:MSR:APPLy
.............................................................................943
[SENSe:]ESPectrum<sb>:MSR:BAND
............................................................................. 943
[SENSe:]ESPectrum<sb>:MSR:BCATegory
...................................................................... 944
[SENSe:]ESPectrum<sb>:MSR:CLASs
............................................................................ 944
[SENSe:]ESPectrum<sb>:MSR:GSM:CARRier
................................................................. 945
[SENSe:]ESPectrum<sb>:MSR:GSM:CPResent
............................................................... 946
[SENSe:]ESPectrum<sb>:MSR:LTE:CPResent
................................................................. 946
[SENSe:]ESPectrum<sb>:MSR:MPOWer
......................................................................... 947
[SENSe:]ESPectrum<sb>:MSR:RFBWidth
........................................................................947
Configuring and Performing Measurements