Remote Commands
R&S
®
FSW
925
User Manual 1173.9411.02 ─ 43
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STOP
<Frequency>
This command defines the stop frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
●
smaller than the span the sweep list covers, the R&S
FSW will not measure the
ranges that are outside the span - results may be invalid.
●
greater than the span the sweep list covers, the R&S
FSW will adjust the start fre-
quency of the first range and the stop frequency of the last range to the span
For more information see
Chapter 7.7, "Spurious Emissions Measurement"
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Frequency>
Numeric value.
*RST:
-2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Default unit: Hz
Example:
LIST:RANG2:STOP 5MHZ
Manual operation:
See
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:ATTenuation
<Attenuation>
This command defines the input attenuation for a SEM range.
In case of high speed measurements, the input attenuation has to be identical for all
ranges.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
<ri>
1..n
Selects the measurement range.
Parameters:
<Attenuation>
Numeric value.
Refer to the data sheet for the attenuation range.
*RST:
10 dB
Default unit: dB
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:ATTenuation:AUTO
<State>
This command turns automatic selection of the input attenuation for a SEM range on
and off.