RF OVLD
Overload of the input mixer or of the analog IF path.
●
Increase the RF attenuation (for RF input).
●
Reduce the input level (for digital input)
UNCAL
One of the following conditions applies:
●
Correction data has been switched off.
●
No correction values are available, for example after a firmware update.
●
Record the correction data by performing a self alignment
(For details refer to
Chapter 5.1.1.6, "Performing a Self-Alignment and a Self-
WRONG_FW
The firmware version is out-of-date and does not support the currently installed hard-
ware. Until the firmware version is updated, this error message is displayed and self-
alignment fails.
(For details refer to
Chapter 12.6.4, "Firmware Updates"
5.4.2
Accessing the Functionality
All tasks necessary to operate the instrument can be performed using this user inter-
face. Apart from instrument specific keys, all other keys that correspond to an external
keyboard (e.g. arrow keys, ENTER key) operate conform to Microsoft.
For most tasks, there are at least 2 alternative methods to perform them:
●
Using the touchscreen
●
Using other elements provided by the front panel, e.g. the keypad, rotary knob, or
arrow and position keys.
The measurement and instrument functions and settings can be accessed by selecting
one of the following elements:
●
System and function keys on the front panel of the instrument
●
Softkeys on the touchscreen
●
Context menus for specific elements on the touchscreen
●
Icons on the tool bar in the touchscreen
●
Displayed setting on the touchscreen
5.4.2.1
Toolbar
Standard functions can be performed via the icons in the toolbar at thetop of the
screen.
You can hide the toolbar display, e.g. when using remote control, to enlarge the display
area for the measurement results ("Setup > Display > Displayed Items"). See the
R&S
FSW User Manual for details.
The following functions are available by default: