General Instrument Setup
R&S
®
FSW
716
User Manual 1173.9411.02 ─ 43
Frequency ← Calibration Frequency RF
Defines the frequency of the internal broadband calibration signal to be used for IF fil-
ter calibration (max. 64
MHz).
For narrowband signals, 64
MHz is sent.
Calibration Frequency MW
Uses the microwave calibration signal as the RF input (for frequencies higher than 8
GHz). This function is used to calibrate the YIG-filter on the microwave converter. The
microwave calibration signal is pulsed.
You can define whether the distance between input pulses is small or wide.
Remote command:
DIAGnostic:SERVice:INPut[:SELect]
DIAGnostic:SERVice:INPut:MC[:DISTance]
DIAGnostic:SERVice:INPut:MC:CFRequency
Calibration Analog Baseband
Uses an internal calibration signal as input to the optional Analog Baseband interface.
This signal is only available if the R&S
FSW-B71 option is installed.
For more information on the Analog Baseband Interface (R&S
FSW-B71) see the
R&S
FSW I/Q Analyzer and I/Q Input User Manual.
Remote command:
DIAGnostic:SERVice:INPut[:SELect]
Calibration Signal Type ← Calibration Analog Baseband
Defines the type of calibration signal to be used for Analog Baseband.
"AC"
1.5625 MHz square wave AC signal
"DC"
DC signal
"DC zero"
no signal
Remote command:
DIAGnostic:SERVice:INPut:AIQ[:TYPE]
12.7.4
Service Functions
Access
: [Setup] > "Service" > "Service Function"
Using service functions
The service functions are not necessary for normal measurement operation. Incorrect
use can affect correct operation and/or data integrity of the R&S
FSW.
Therefore, only user accounts with administrator rights can use service functions and
many of the functions can only be used after entering a password. These functions are
described in the instrument service manual.
Service Functions
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...