To activate a key, select the key on the touchscreen.
To simulate the use of the rotary knob, use the additional keys displayed between the
keypad and the arrow keys:
Icon
Function
Turn left
Enter
Turn right
Mini front panel
The mini front panel provides only the keys on the touchscreen, to operate the
R&S
FSW via an external monitor or remote desktop.
By default, the "Auto close" option is activated and the mini front panel window closes
automatically after you select a key. This is useful if you only require the mini front
panel display occasionally to press a single function key.
If you want the window to remain open, deactivate the "Auto close" option. You can
close the window manually by selecting "Close planel" or the key combination [ALT +
M] (be aware of the keyboard language defined in the operating system!).
To display the soft front panel or mini front panel
1. Press the [Setup] key and select the "Display" softkey.
2. Select the "Displayed Items" tab.
3. Select "Front Panel": "On" or "Mini Front Panel": "On".
To activate or deactivate the front panel temporarily, press the [F6] key on the external
keyboard (if available) or on the remote computer.
12.3
Transducers
12.3.1
Basics on Transducer Factors
The transducer allows you to manipulate the trace at discrete trace points to correct
the signal coming from an input device. Transducers are often used to correct the fre-
quency response for antennas, for example. The transducer is configured by defining
transducer factors for specific trace points. A set of transducer factors defines an inter-
polated transducer line and can be stored on the instrument.
In the Spectrum application, the correction factor from all active transducers is calcula-
ted for each displayed trace point once in advance and is added to the result of the
level measurement during the sweep. If the sweep range changes, the correction val-