Common Measurement Settings
R&S
®
FSW
447
User Manual 1173.9411.02 ─ 43
this case, the signal trace is surveyed in a specified bandwidth around the expected
center frequency. After each sweep, the center frequency is set to the maximum signal
found within the searched bandwidth. If no maximum signal above a defined threshold
value is found in the searched bandwidth, the center frequency remains unchanged.
The search bandwidth and the threshold value are shown in the diagram by red lines
which are labeled as "TRK".
Signal Tracking
Access
: "Overview" > "Frequency" > "Signal Tracking" tab
Defines the settings for signal tracking. These settings are only available for spans > 0.
For more details see
Chapter 8.3.3, "Keeping the Center Frequency Stable - Signal
If activated, after each sweep, the center frequency is set to the maximum level of the
specified "Signal Track Trace" found within the searched "Tracking Bandwidth".
If the signal level does not pass the "Tracking Threshold", the center frequency is not
changed.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:STRack[:STATe]
CALCulate<n>:MARKer<m>:FUNCtion:STRack:BANDwidth
CALCulate<n>:MARKer<m>:FUNCtion:STRack:THReshold
CALCulate<n>:MARKer<m>:FUNCtion:STRack:TRACe
8.3.4
How To Define the Frequency Range
The following step-by-step instructions demonstrate how to configure the frequency
and span settings. For details on individual functions and settings see
Frequency and Span Configuration