Common Measurement Settings
R&S
®
FSW
436
User Manual 1173.9411.02 ─ 43
c) From the
C:\R_S\INSTR\USER\cvl
directory, select the
100112_MAG_4_B5000_3G5.B5G
file.
d) Select [ENTER].
3. Switch to the "Mixer Settings" tab.
4. Select the "Mixer Type": "3-port".
5. Select "Mixer Settings > Band" to define the required frequency range.
6. From the "Band" selection list, select the band "V".
7. A span is automatically set which covers the whole V band (50
GHz to 75
GHz).
8. In the "Mixer Settings", select "Conversion Loss: Table" for Range
1 to define fre-
quency-dependent level correction.
9. From the selection list, select the
100112_MAG_4_B5000_3G5.B5G
conversion
loss table which you imported previously. No further settings are necessary since
the selected file contains all required parameters.
Now you can analyze the broadband signal with an analysis bandwidth of 5
GHz.
8.2.6
Output Settings
Access
: [Input/Output] > "Output"
The R&S
FSW can provide output to special connectors for other devices.
For details on connectors, refer to the R&S
FSW Getting Started manual, "Front / Rear
Panel View" chapters.
Providing trigger signals as output is described in
Chapter 8.2.7, "Trigger Input/Output
Providing output for LISN control in EMI measurements is described in
ter 7.13.4.3, "LISN Control Settings"
Data Input and Output