Measurements and Results
R&S
®
FSW
230
User Manual 1173.9411.02 ─ 43
Frequency
Defines the frequency of the signal provided by the signal generator.
If the
Generator Frequency Coupling State
is active, this value is read-only and indi-
cates the resulting frequency setting on the generator.
Remote command:
CONFigure:GENerator:FREQuency:CENTer
CONFigure:GENerator:NPRatio:FREQuency:CENTer:CSTate?
Level (RMS)
(Default:) The specified power level is used for the output power by the connected sig-
nal generator.
Remote command:
CONFigure:GENerator:POWer:LEVel
CONFigure:GENerator:NPRatio:POWer:LEVel:CSTate?
Level Offset
Defines a fixed offset in the power level used by the generator, for example due to a
gain from the DUT.
Remote command:
CONFigure:GENerator:POWer:LEVel:OFFSet
CONFigure:GENerator:NPRatio:POWer:LEVel:OFFSet:CSTate?
Reference Frequency
Selects the source of the generator reference frequency.
The internal reference is that of the signal generator itself.
An external reference is provided via the EXT connectors on the generator, for exam-
ple by the R&S
FSW.
Remote command:
CONFigure:GENerator:EXTernal:ROSCillator
CONFigure:GENerator:NPRatio:EXTernal:ROSCillator:CSTate?
Standard
For reference only: the standard currently used by the signal generator. If no standard
is defined on the generator, notched signals are not available (see
Remote command:
CONFigure:GENerator:NPRatio:BB:STANdard?
ARB Waveform File
For reference only: the ARB waveform file currently used by the signal generator.
Remote command:
CONFigure:GENerator:NPRatio:BB:ARBitrary:WAVeform:SELect?