BIDir
Changes of a parameter are applied both ways (from channel 1
to channel 2 and vice versa).
LTOR
Changes of a parameter are applied from channel 1 to channel
2, but not the other way around.
RTOL
Changes of a parameter are applied from channel 2 to channel
1, but not the other way around.
Example:
INST:COUP:USER:REL BID
Selects bidirectional changes for the user-defined coupling with
the index number 1.
Manual operation:
See
INSTrument:COUPle:USER<uc>:REMove
[<Scope>]
This command deletes a user-defined coupling mechanism.
Suffix:
<uc>
.
Index of a user-defined parameter coupling. Which indexes are
available is defined by
Parameters:
<Scope>
ALL
Optional SCPI parameter, used instead of the <uc> suffix.
Deletes all user-defined couplings.
Example:
INST:COUP:USER3:REM
Removes the user-defined coupling with the index number 3.
Manual operation:
See
See
INSTrument:COUPle:USER<uc>:STATe
<State>
Enables or disables the specified user-defined parameter coupling.
Suffix:
<uc>
.
Index of a user-defined parameter coupling. Which indexes are
available is defined by
Parameters:
<State>
ON | OFF | 1 | 0
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0