Remote Commands
R&S
®
FSW
1312
User Manual 1173.9411.02 ─ 43
INSTrument:COUPle:ACDC
<State>
This command turns synchronization of the AC / DC Coupling state between measure-
ment channels on and off.
Parameters:
<State>
ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
ALL
Example:
INST:COUP:ACDC ALL
Synchronizes the "AC/DC Coupling" parameter.
Manual operation:
See
"Synchronizing parameters across all measurement chan-
INSTrument:COUPle:ATTen
<State>
This command turns synchronization of the attenuation and unit between measure-
ment channels on and off.
Parameters:
<State>
ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
ALL
Example:
INST:COUP:ATT ALL
Synchronizes the attenuation.
Manual operation:
See
"Synchronizing parameters across all measurement chan-
INSTrument:COUPle:AUNit
<State>
This command turns synchronization of the amplitude unit configuration between mea-
surement channels on and off.
Parameters:
<State>
ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
NONE
Example:
INST:COUP:AUN ALL
Configuring the R&S