Remote Commands
R&S
®
FSW
1271
User Manual 1173.9411.02 ─ 43
String
Unit
'DBUV/M'
dB
μ
V/m
(Requires R&S
FSW-K54 (EMI measurements) option.)
'DBUA'
dB
μ
A
'DBUA/M'
dB
μ
A/m
(Requires R&S
FSW-K54 (EMI measurements) option.)
'DBPW'
dBpW
'DBPT'
dBpT
MMEMory:LOAD<n>:TFACtor
<FileName>
Loads the transducer factor from the selected file in
.CSV
format.
Suffix:
<n>
.
irrelevant
Parameters:
<FileName>
String containing the path and name of the
CSV
import file.
Example:
MMEM:LOAD:TFAC 'C:\TEST.CSV'
Manual operation:
See
MMEMory:STORe<n>:TFACtor
<FileName>, <TransdName>
This command exports transducer factor data to an ASCII (CSV) file.
For details on the file format see
Chapter 12.3.3, "Reference: Transducer Factor File
Suffix:
<n>
.
irrelevant
Parameters:
<FileName>
Name of the transducer factor to be exported.
<TransdName>
Name of the transducer factor to be exported.
Example:
MMEM:STOR:TFAC 'C:\TEST', 'Transducer1'
Stores the transducer factor named "Transducer1" in the file
TEST.CSV
.
Manual operation:
See
14.10.4
Compensating for Frequency Response Using Touchstone files
(R&S
FSW-K544)
If the Frequency Response Correction option (R&S
FSW-K544) is installed, the
R&S
FSW supports the use of Touchstone (
.SnP
) files, as well as additional frequency
response correction files in
.fres
format.