Remote Commands
R&S
®
FSW
1131
User Manual 1173.9411.02 ─ 43
RMS
RMS value
AVERage
Average
CAVerage
CISPR Average (CISPR filter only)
CRMS
CISPR RMS (CISPR filter only)
*RST:
APEak
Example:
DET POS
Sets the detector to "positive peak".
Manual operation:
See
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO
<State>
This command couples and decouples the detector to the trace mode.
Suffix:
<n>
.
<t>
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
DET:AUTO OFF
The selection of the detector is not coupled to the trace mode.
Manual operation:
See
TRACe<n>:COPY
<TraceNumber>, <TraceNumber>
This command copies data from one trace to another.
Suffix:
<n>
.
Parameters:
<TraceNumber>
TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6
The first parameter is the destination trace, the second parame-
ter is the source.
(Note the 'e' in the parameter is required!)
Example:
TRAC:COPY TRACE1,TRACE2
Copies the data from trace 2 to trace 1.
Manual operation:
See
Analyzing Measurements (Basics)