Remote Commands
R&S
®
FSW
1042
User Manual 1173.9411.02 ─ 43
Parameters:
<Attenuation>
Range:
see data sheet
Increment: 5 dB (with optional electr. attenuator: 1 dB)
*RST:
10 dB (AUTO is set to ON)
Default unit: DB
Example:
INP:ATT 30dB
Defines a 30 dB attenuation and decouples the attenuation from
the reference level.
Manual operation:
See
INPut<ip>:ATTenuation:AUTO
<State>
This command couples or decouples the attenuation to the reference level. Thus, when
the reference level is changed, the R&S
FSW determines the signal level for optimal
internal data processing and sets the required attenuation accordingly.
Suffix:
<ip>
.
1 | 2
For R&S
FSW85 models with two RF input connectors:
1: Input 1 (1
mm [RF Input] connector)
2: Input 2 (1.85
mm [RF2 Input] connector)
For all other models:
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
INP:ATT:AUTO ON
Couples the attenuation to the reference level.
Manual operation:
See
INPut<ip>:ATTenuation:AUTO:MODE
<OptMode>
Selects the priority for signal processing
after
the RF attenuation has been applied.
Suffix:
<ip>
.
1 | 2
For R&S
FSW85 models with two RF input connectors:
1: Input 1 (1
mm [RF Input] connector)
2: Input 2 (1.85
mm [RF2 Input] connector)
For all other models:
irrelevant
Parameters:
<OptMode>
LNOise | LDIStortion
LNOise
Optimized for high sensitivity and low noise levels
LDIStortion
Optimized for low distortion by avoiding intermodulation
Setting Basic Measurement Parameters