Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
246
User Manual 1173.9240.02 ─ 23
Example:
INP:ATT 30dB
Defines a 30 dB attenuation and decouples the attenuation from
the reference level.
Manual operation:
See
INPut:ATTenuation:AUTO
<State>
This command couples or decouples the attenuation to the reference level. Thus, when
the reference level is changed, the R&S
FSW determines the signal level for optimal
internal data processing and sets the required attenuation accordingly.
This function is not available if the optional Digital Baseband Interface is active.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
INP:ATT:AUTO ON
Couples the attenuation to the reference level.
Manual operation:
See
INPut:ATTenuation:AUTO:MODE
<OptMode>
Selects the priority for signal processing
after
the RF attenuation has been applied.
Parameters:
<OptMode>
LNOise | LDIStortion
LNOise
Optimized for high sensitivity and low noise levels
LDIStortion
Optimized for low distortion by avoiding intermodulation
*RST:
LDIStortion
Example:
INP:ATT:AUTO:MODE LNO
INPut:EATT
<Attenuation>
This command defines an electronic attenuation manually. Automatic mode must be
switched off (
INP:EATT:AUTO OFF
If the current reference level is not compatible with an attenuation that has been set
manually, the command also adjusts the reference level.
This command requires the electronic attenuation hardware option.
It is not available if the optional Digital Baseband Interface is active.
Configuring the Measurement