Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
763
User Manual 1178.8520.02 ─ 01
CALCulate<n>:LIMit<li>:ACPower:ACHannel[:RELative]:STATe
This command turns the relative limit check for the adjacent channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe]
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Manual operation:
See
See
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:ABSolute
<LowerLimit>[,
<UpperLimit>]
This command defines the absolute limit of the alternate channels.
If you have defined an absolute limit as well as a relative limit, the R&S
FSV/A uses the
lower value for the limit check.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<ch>
1..n
Alternate channel number
Parameters:
<LowerLimit>
The limit of the lower adjacent channel.
Range:
-200 dBm to 200 dBm
*RST:
-200 dBm
Default unit: dBm
<UpperLimit>
The limit of the upper adjacent channel.
Range:
-200 dBm to 200 dBm
*RST:
-200 dBm
Default unit: dBm
Manual operation:
See
See
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:ABSolute:STATe
This command turns the absolute limit check for the alternate channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe]
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Configuring and Performing Measurements