General Instrument Setup
R&S
®
FSVA3000/ R&S
®
FSV3000
611
User Manual 1178.8520.02 ─ 01
Filename ← Frequency Response File Information
Name of a loaded Touchstone file.
Magnitude ← Frequency Response File Information
Activates or deactivates the use of the correction data in the selected file for magnitude
results.
Remote command:
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:MAGNitude[:STATe]
Phase ← Frequency Response File Information
Activates or deactivates the use of the correction data in the selected file for phase
results.
Remote command:
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:PHASe[:STATe]
Add Freq Resp File
Loads a frequency response (
.fres
) file to the current configuration. The maximum
number of files per configuration is 15.
The default directory for
.fres
files is
C:\R_S\INSTR\USER\Fresponse
.
Remote command:
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:INSert
Remove Frequency Response File
Removes the selected frequency response (
.fres
) file from the current configuration.
Remote command:
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:REMove
[SENSe:]CORRection:FRESponse:USER:FLISt<fli>:CLEar
Preview
The preview of the (combined) user correction files shows the correction values.
Remote command:
[SENSe:]CORRection:FRESponse:USER:PSTate
Selected File ← Preview
The preview of the selected user correction file shows the correction values for the
specified frequency range. The values for individual ports can be activated or deactiva-
ted.
Frequency Response Correction (R&S
FSV/A-K544)