Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
225
User Manual 1178.8520.02 ─ 01
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:TRANsducer
Limit Check <n>
Sets the type of limit check for the n-th power class in the range. Up to four limits are
possible.
For details on limit checks, see
Chapter 7.6.4.2, "Limit Lines in SEM Measurements"
The limit state affects the availability of all limit settings.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:STATe
Abs Limit Start / Stop <n>
Sets an absolute limit value for the n-th power class at the start or stop frequency of
the range [dBm].
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:ABSolute:STARt
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:ABSolute:STOP
Rel Limit Start / Stop <n>
Sets a relative limit value for the n-th power class at the start or stop frequency of the
range [dBc].
By default, this value is a fixed relative limit, i.e. no function is defined. To define a
function for the relative limit, select the input field for "Rel Limit Start" or "Rel Limit
Stop" and then the "f(x)" icon that appears.
If the function is set to "Max" , you can define a relative
and
an absolute limit level. In
this case, the maximum of the two values is used as the limit level.
For more information, see
"Relative limit line functions"
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP
Spectrum Emission Mask (SEM) Measurement