Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
197
User Manual 1178.8520.02 ─ 01
Remote command:
You can also query the determined carrier-to-noise ratio via the remote command
CALC:MARK:FUNC:POW:RES? CN
or
CALC:MARK:FUNC:POW:RES? CN0
, see
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
7.4.3
Carrier-to-Noise Configuration
Access
: "Overview" > "Select Measurement" > "C/N" / "C/N0" > "Carrier Noise Config"
Both a carrier-to-noise ratio (C/N) and a carrier-to-noise ratio in relation to the band-
width (C/N
0
) measurement are available.
Carrier-to-noise measurements are not available in zero span mode.
The easiest way to configure a measurement is using the configuration "Overview" ,
see
Chapter 8.1, "Configuration Overview"
Auto measurement
Use the
function to determine the required measurement parameters automat-
ically.
The following settings are determined:
●
Center frequency
●
Channel bandwidth: determined signal bandwidth * 2
See also
Chapter 7.1, "Auto Measurements"
The remote commands required to perform these tasks are described in
ter 14.5.4, "Measuring the Carrier-to-Noise Ratio"
Carrier-to-Noise Measurements