Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
175
User Manual 1178.8520.02 ─ 01
Remote command:
Activating/Deactivating:
[SENSe:]POWer:ACHannel:FILTer[:STATe]:SBLock<sb>:CHANnel<ch>
on page 768
Alpha value:
[SENSe:]POWer:ACHannel:FILTer:ALPHa:SBLock<sb>:CHANnel<ch>
7.3.5.3
MSR Adjacent Channel Setup
Access
: "Overview" > "Select Measurement" > "Channel Power ACLR" > "CP / ACLR
Standard" > "Standard" : "Multi-Standard Radio" > "CP / ACLR Config" > "Adjacent
Channels" tab
The "Adjacent Channels" tab provides all the channel settings to configure adjacent
and gap channels in MSR ACLR measurements.
For symmetrical channel definition (see
" Symmetrical Adjacent Setup "
the dialog box is reduced as the upper and lower channels are identical.
Figure 7-28: Asymmetrical adjacent channel definition
Channel Power and Adjacent-Channel Power (ACLR) Measurement