Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1141
User Manual 1178.8520.02 ─ 01
INSTrument:COUPle:PRESel
<State>
This command turns synchronization of the preselector state between measurement
channels on and off.
Parameters:
<State>
ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
ALL
Example:
INST:COUP:PRES ALL
Synchronizes the preselector configuration.
INSTrument:COUPle:RLEVel
<State>
This command turns synchronization of the reference level between measurement
channels on and off.
Parameters:
<State>
ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
NONE
Example:
INST:COUP:RLEV ALL
INSTrument:COUPle:SPAN
<State>
This command turns synchronization of the start and stop frequency between mea-
surement channels on and off.
Parameters:
<State>
ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
NONE
Example:
INST:COUP:SPAN ALL
Synchronizes the start and stop frequency.
Configuring the R&S