Instrument Tour
R&S
®
FSVA3000/R&S
®
FSV3000
36
Getting Started 1330.8073.02 ─ 04
5.1.1
Power Key
The power key is on the lower left corner of the front
panel. It starts up and shuts down the instrument.
5.1.2
USB
The front panel provides three female USB connectors (USB-A) to connect devi-
ces like a keyboard or a mouse. In addition, a memory stick can be connected to
store and reload instrument settings and measurement data.
The USB connectors support standard 2.0.
The rear panel provides further USB connectors, including some that sup-
port USB standard 3.0.
5.1.3
System Keys
System keys set the instrument to a predefined state, change basic settings, and
provide print and display functions.
A detailed description of the corresponding functions is provided in the user man-
ual.
Table 5-1: System keys
System key
Assigned functions
[Preset]
Resets the instrument to the default state.
[Setup]
Provides basic instrument configuration functions, e.g.:
●
Reference frequency (external/internal), noise source
●
Date, time, display configuration
●
LAN interface
●
Self-alignment
●
Firmware update and enabling of options
●
Information about instrument configuration incl. firmware ver-
sion and System error messages
●
Service support functions (self-test etc.)
[Print]
Provides print and screenshot functions