Analysis
R&S
®
FSV/A3000 I/Q
Analyzer
95
User Manual 1178.8536.02 ─ 01
For details see the R&S
FSV/A User Manual.
Time Stamp
Activates and deactivates the timestamp. The timestamp shows the system time while
the measurement is running. In single sweep mode or if the sweep is stopped, the
timestamp shows the time and date of the end of the sweep.
When active, the timestamp replaces the display of the frame number.
Remote command:
CALCulate<n>:SPECtrogram:TSTamp[:STATe]
CALCulate<n>:SPECtrogram:TSTamp:DATA?
Color Mapping
Opens the "Color Mapping" dialog.
For details see the R&S
FSV/A User Manual.
Continuous Sweep / Run Cont
After triggering, starts the sweep and repeats it continuously until stopped. This is the
default setting.
After triggering, starts the measurement and repeats it continuously until stopped.
While the measurement is running, the "Continuous Sweep" softkey and the [RUN
CONT] key are highlighted. The running measurement can be aborted by selecting the
highlighted softkey or key again. The results are not deleted until a new measurement
is started.
Note:
Sequencer. If the Sequencer is active, the "Continuous Sweep" softkey only con-
trols the sweep mode for the currently selected channel. However, the sweep mode
only takes effect the next time the Sequencer activates that channel, and only for a
channel-defined sequence. In this case, a channel in continuous sweep mode is swept
repeatedly.
If the Sequencer is active in MSRT mode, the "Continuous Sweep" function does not
start data capturing. It merely affects trace averaging over multiple sequences. In this
case, trace averaging is performed.
Furthermore, the [RUN CONT] key controls the Sequencer, not individual sweeps.
[RUN CONT] starts the Sequencer in continuous mode.
Remote command:
Single Sweep / Run Single
After triggering, starts the number of sweeps set in "Sweep Count". The measurement
stops after the defined number of sweeps has been performed.
While the measurement is running, the "Single Sweep" softkey and the [RUN SINGLE]
key are highlighted. The running measurement can be aborted by selecting the high-
lighted softkey or key again.
Note:
Sequencer. If the Sequencer is active, the "Single Sweep" softkey only controls
the sweep mode for the currently selected channel. However, the sweep mode only
takes effect the next time the Sequencer activates that channel, and only for a chan-
nel-defined sequence. In this case, the Sequencer sweeps a channel in single sweep
mode only once.
Spectrogram Settings