Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
234
User Manual 1178.8536.02 ─ 01
If necessary, the command activates the delta marker and positions a reference
marker to the peak power.
Suffix:
<n>
.
<m>
Parameters:
<Position>
Numeric value that defines the marker position on the x-axis.
Range:
The value range and unit depend on the measure-
ment and scale of the x-axis.
Example:
CALC:DELT:X?
Outputs the absolute x-value of delta marker 1.
Manual operation:
See
CALCulate<n>:MARKer<m>:AOFF
This command turns off all markers.
Suffix:
<n>
.
<m>
Example:
CALC:MARK:AOFF
Switches off all markers.
Manual operation:
See
CALCulate<n>:MARKer<ms>:LINK:TO:MARKer<md>
<State>
This command links normal marker <m1> to any active normal marker <m2>.
If you change the horizontal position of marker <m2>, marker <m1> changes its hori-
zontal position to the same value.
Suffix:
<n>
.
<ms>
source marker, see
<md>
destination marker, see
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK4:LINK:TO:MARK2 ON
Links marker 4 to marker 2.
Manual operation:
See
I/Q Analysis