Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
141
User Manual 1178.8536.02 ─ 01
OFF | 0
The analog mixer path is always used.
Example:
INP:DPAT OFF
Manual operation:
See
INPut<ip>:FILTer:YIG[:STATe]
<State>
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1 (0 for I/Q Analyzer, GSM, VSA, Pulse, Amplifier,
Transient Analysis, DOCSIS and MC Group Delay
measurements)
Example:
INP:FILT:YIG OFF
Deactivates the YIG-preselector.
Manual operation:
See
INPut<ip>:IMPedance
<Impedance>
This command selects the nominal input impedance of the RF input. In some applica-
tions, only 50
Ω
are supported.
The command is not available for measurements with the optional Digital Baseband
Interface.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Impedance>
50 | 75
numeric value
User-defined impedance from 50 Ohm to 100000000 Ohm
(=100 MOhm)
User-defined values are only available for the Spectrum applica-
tion, the I/Q Analyzer, and some optional applications.
*RST:
50 Ω
Default unit: OHM
Example:
INP:IMP 75
Manual operation:
See
See
INPut<ip>:IMPedance:PTYPe
<PadType>
Defines the type of matching pad used for impedance conversion for RF input.
Configuring I/Q Analyzer Measurements