Remote Commands for Bluetooth Measurements (K8)
R&S
®
FSV-K8
74
Operating Manual 1176.7610.02 ─ 02
Manual operation:
See
CALCulate<n>:BTOoth:CFSTability:FERRor:BLOCk?
<Mode>
This command reads the maximum frequency deviation of all blocks in the Carrier Fre-
quency Stability and Modulation Accuracy measurement.
Note:
This command is only available with active Carrier Frequency Stability and Modulation
Accuracy measurement (see
With all other measurements this command causes a query error.
Suffix:
<n>
.
irrelevant
Parameters:
<Mode>
MINimum | MAXimum | AVERage
*RST:
-
Example:
For the first steps refer to
chapter 4.14, "Programming Exam-
CONF:BTO:MEAS CFST
Activates the Carrier Frequency Stability and Modulation Accu-
racy measurement.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:CFST:FERR:BLOC? AVER
Reads the maximum average frequency deviation of all blocks.
Usage:
Query only
Mode:
BT
Manual operation:
See
CALCulate<n>:BTOoth:CFSTability:FERRor:INITial?
<Mode>
This command reads the combined frequency deviation of all packets and all blocks in
the Carrier Frequency Stability and Modulation Accuracy measurement.
Note:
This command is only available with active Carrier Frequency Stability and Modulation
Accuracy measurement (see
With all other measurements this command causes a query error.
Suffix:
<n>
.
irrelevant
Parameters:
<Mode>
MINimum | MAXimum | AVERage
*RST:
-
CALCulate subsystem