Instrument Functions for Bluetooth Measurements (K8)
R&S
®
FSV-K8
40
Operating Manual 1176.7610.02 ─ 02
Search Len Manual ← Find Sync ← Output Power
Opens an edit dialog box to enter the search length used for determining the sync word
or the burst. The unit of the search length is seconds; the values range from 100
μ
s to
(400001 / points per symbol)
μ
s. The default setting is 1825
μ
s.
The selected number of points per symbol and the maximum search length is listed in
the table below.
Points per symbol
Max. search length
(number of slots)
2
104.4
4
52.2
8
26.1
16
13.1
32
6.5
For information on the correlation of trigger and record length refer to the
Remote command:
Power Avg Start (Output Power) ← Output Power
Opens an edit dialog box to enter the start position of the evaluation area for the aver-
age burst power. The values range is from 0 to 100%, the default setting is 20%.
For further information refer to
chapter 2, "Introduction to Bluetooth Measurements
Remote command:
Power Avg Stop (Output Power) ← Output Power
Opens an edit dialog box to enter the stop position of the evaluation area for the aver-
age burst power. The values range from 0 to 100%, the default setting is 80%.
For further information refer to
chapter 2, "Introduction to Bluetooth Measurements
Remote command:
TX Spec ACLR
Opens a submenu to configure the TX Spec ACLR measurement. For further details
refer to
chapter 3.1.2, "Adjacent Channel Power Measurement"
Remote command:
CONF:BTO:MEAS ACLR
, see
Softkeys of the Bluetooth Menu