Remote Commands for Bluetooth Measurements (K8)
R&S
®
FSV-K8
116
Operating Manual 1176.7610.02 ─ 02
[SENSe:]SWEep:COUNt
<NumberSweeps>
This command defines the number of sweeps started with single sweep, which are
used for calculating the average or maximum value. If the values 0 or 1 are set, one
sweep is performed.
Parameters:
<NumberSweeps>
0 to 32767
*RST:
0 (GSM: 200, PHN:1)
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
[SENSe:]SWEep:COUNt:CURRent?
This query command returns the current number of started sweeps. A sweep count
value should be set and the device should be in single sweep mode.
Example:
SWE:COUNt 64
Sets sweep count to 64
INIT:CONT OFF
Switches to single sweep mode
INIT
Starts a sweep (without waiting for the sweep end!)
SWE:COUN:CURR?
Queries the number of started sweeps
Usage:
Query only
[SENSe:]SWEep:EGATe:HOLDoff
<DelayTime>
This command defines the delay time between the external gate signal and the contin-
uation of the sweep.
Note
: Using gate mode "level" (see
and an IFP trigger (see
on page 126), the hold-
off time for the IFP trigger is ignored for frequency sweep, FFT sweep, zero span and
IQ mode measurements.
Parameters:
<DelayTime>
0 s to 30 s
*RST:
0s
Example:
SWE:EGAT:HOLD 100us
Manual operation:
See
"Gate Delay (Spurious Emissions)"
SENSe Subsystem (BLUETOOTH, K8)