Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
84
Operating Manual 1176.7578.02 ─ 09
Example:
Figure 3-11: Result display for "Symbols" in binary mode
If a pattern search is active, a found pattern is indicated in the symbol table, as well.
SCPI commands:
CALC:FEED 'XTIM:DDEM:SYMB'
to define the required source type (see
TRAC:DATA TRACE1
to query the trace results (see
3.1.1.17
Error Vector Magnitude (EVM)
Displays the error vector magnitude as a function of symbols or time.
C
t
EV
t
EVM
with t=n
·
T
D
and T
D
=the duration of one sampling period at the sample rate defined by
the display points per symbol parameter (see
The normalization constant C is chosen according to the EVM normalization. By
default C² is the mean power of the reference signal.
k
T
k
REF
K
C
2
1
and
periods
symbol
of
duration
T
Note that k=0.5
·
n
·
T for Offset QPSK with inactive Offset EVM.