Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
331
Operating Manual 1176.7578.02 ─ 09
Example:
TRIG:SLOP NEG
Manual operation:
See
TRIGger<n>[:SEQuence]:SOURce
<Source>
This command selects the trigger source.
IF power and RF power triggers are not available together with the bandwidth exten-
sion option R&S
FSV-B160.
For details on trigger modes refer to the "Trg/Gate Source" softkey in the base unit
description.
Suffix:
<n>
.
irrelevant
Parameters:
<Source>
IMMediate
Free Run
EXTern
External trigger
IFPower
Power trigger at the second intermediate frequency
BBPower
Baseband power (for digital input via the R&S Digital I/Q Inter-
face, R&S
FSV-B17)
*RST:
IMMediate
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Manual operation:
See
See
See
See
TRIGger subsystem